URAIAN SINGKAT PEKERJAAN
1. Lingkup 1.1 Kode RUP : 51523432
Pekerjaan
1.2 Nama paket pengadaan : Pengadaan Kontrak Service
Maintenance Alat X-Ray C Arm Merk GE di Instalasi
Radiodiagnostik Pusat Kanker Nasional Rumah Sakit Kanker
“Dharmais” Tahun 2024
1.3 Uraian singkat paket pengadaan :
Kontrak service vital++
Nama alat : X-Ray C-Arm
Merk : GE
Type : 9900 0" STD C ARM
SN : 7982XR0008
Lingkup pekerjaan :
1. Melaksanakan Preventive Maintenance (PM) : 2x/tahun
Alat X-Ray C-Arm :
System external inspection :
-. System labels
-. AC power cord and plug
-. Interconnect cable & LEMO connector
-. High voltage (HV) cable inpection
-. Antistatic drag wires
-. Shelf fit check (9800)
-. Monitor arm
-. Workstation brake and wheels
-. C-Arm wheels, steering & brakes
-. C-Arm mechanical movement
-. Skin spacer
-. Ground continuity
-. Network connector
-. System covers
Internal inspection :
-. Air vents and filters
-. C-Arm battery packs
-. X-Ray tube inspection
-. HV cable connector/receptable cleaning & greasing
Function checks :
-. Power on checks
-. UPS battery checks
-. Cooling fans operation
-. Laser aimer
-. X-Ray keyswitch/footswitch/handleswitch
-. X-Ray indicators
-. C-Arm controls
-. Reference monitor
-. Motorized movement
-. DC power supply PS1 calibration
-. C-Arm batteries & charger
-. Touchscreen accurancy
-. Workstation checks
-. Isolation transformer output
-. Software version, hour meter reading & unique ID
-. Printer
Imaging checks kVP accurancy check :
-. Flouro 60 kVp
-. Flouro 90 kVp
-. Flouro 120 kVp
-. Radiographic 60 kVp @50ma
-. Radiographic 90 kVp @50ma
-. Radiographic 120 kVp @50ma
-. Radiographic 60 kVp @200ma
-. Radiographic 90 kVp @200ma
-. Radiographic 100 kVp @200ma
mA accurancy check :
-. 60 kVp & 10 mA, TP72 & TP73
-. 90 kVp & 10 mA, TP72 & TP73
-. 60 kVp & 5 mA, TP61 & TP63
-. 90 kVp & 5 mA, TP61 & TP63
Filament calibration :
-. Filament calibration
High voltage arc test :
-. High voltage arc test
Beam alignment verification :
-. Beam alignment verification
Half value layer check :
-. Half value layer check
Image receptor entrance dose rate check :
-. Standard fluoro
-. High level fluoro
Reference AKR limit checks :
-. Reference AKR limit for standard fluoro
-. Reference AKR limit for high level fluoro (HLF)
Dose display accuracy check :
-. Calculated DAP percent error
-. Calculated CAK percent error
-. Calculated AKR percent error
Auto technigue tracking :
-. Auto technigue tracking
Image solution check :
-. Normal
-. Mag 1
-. Mag 2
System backup & cleanup :
-. Anode rotor test
-. Inerlock test
-. Intelligent shutdown
2. Jasa service & troubleshooting
3. Layanan on call setiap hari kerja bila terjadi
kerusakan (remote service)
4. Waktu respon dalam 1 jam setelah laporan
diterima
Hari kerja : Senin - Jum'at, 09.00 - 17.00
Hari Sabtu : 09.00 - 14.00
5. Kalibrasi oleh lembaga kalibrasi yang
bersertifikasi KAN 1 kali setahun
6. Termasuk standard part coverage bila terjadi
kerusakan
7. Uptime Guarantee : 80%
8 Harga sudah termasuk PPN 11%
1.4 Jangka waktu penyelesaian pekerjaan : sampai dengan
Desember 2024 ).
1.5 Nama Satuan Kerja/Perangkat daerah : Pusat Kanker
Nasional Rumah Sakit Kanker “Dharmais”
1.6 Nama UKPBJ: Unit Pengadaan Barang/Jasa Pusat Kanker
Nasional Rumah Sakit Kanker “Dharmais”
1.7 Nama Pokja Pemilihan: Pokja Pemilihan Alat Kedokteran,
Barang Non Medik dan Konstruksi
1.8 Alamat Pokja Pemilihan: Jl. Letjen S. Parman Kav. 84-86,
Slipi, Jakarta Barat 11420
1.9 Website Satuan Kerja/Kementerian/Lembaga/ Perangkat
Daerah : www.dharmais.go.id
1.10 Website Aplikasi SPSE :
http://www.lpse.kemkes.go.id/eproc4/