Lampiran Uraian Pekerjaan
Pekerjaan Kontrak Service Pemeliharaan Unit Alat X Ray Flouroscopy Luminos Fusion VE FD Di Instalasi Radiologi
RS Adam Malik Tahun Anggaran 2025
No. Nama Alat dan Uraian Pekerjaan Volume Merek
Siemens-
1 Maintenance dan Kontrak Service Unit Fluoroscopy LUMINOS Fusion VE FD: Type Performance TOP 1 Unit
Luminos Fusion
VE FD SN.
Specifications :
31203
● Suku cadang Normal : Semua suku cadang yang ada pada unit Luminos Fusion VE FD
: (Tidak Termasuk X-Ray Tube, Detector dan Aksesoris)
● Preventive Maintenance (PM) : 2 (dua) kali per tahun
● Corrective Maintenance (CM) : Tidak terbatas (pada saat alat bermasalah atau mengalami
kerusakan)
● Tenaga kerja dan perjalanan CM : Termasuk
● Tenaga kerja dan perjalanan PM : Termasuk
● tenaga kerja jarak jauh : Termasuk
● Dukungan telepon teknis : Termasuk
● Waktu respons awal : 2 (dua) jam
● Waktu respon di tempat (on-site) : Hari kerja berikutnya
● Waktu pengiriman suku cadang : 14 (empat belas) hari kerja
● Layanan jarak jauh teknisi (remote) : Remote pro-active support include
● Safety dan Update Software : Termasuk (jika ada instruksi dari pabrik)
Jendela Pemeliharaan Perbaikan
● : Setiap hari kerja dan jam kerja
(CM
Jendela Pemeliharaan
● : Setiap hari kerja dan jam kerja
Direncanakan (PM)
● Kalibrasi oleh instansi kalibrasi : Tidak Termasuk
yang berwenang dilengkapi
dengan sertifikat, hasil uji serta
label kalibras
Preventive Maintenance Rutin, meliputi :
I General Visual Check : Visual Check per IEC 62353
: Checking the cover panels
: Required operator documents present
: Checking the Cables, Corrugated Hoses
: Check the cable corrugated hoses (of mobile components)
: Check the radiation protective panels
: Check warning labels
II Cleaning the component : Inpection of internal heat dissipation and cleaning
III Polydoros : : Inspection and Maintance
: Checking the external Voltage
: Checking The shelding on the HV Cables
: Radiation Indicator : Radiation Light Check
: Performing Tube adjustment Conditioning The X-ray Tube (s)
IV Luminos Fusion VE FD : : Check of the Basic Unit for Damage
: Evaluating the error log
: Check the cables and corrugated hoses
: Check the chain segment and tilt drive
No. Nama Alat dan Uraian Pekerjaan Volume Merek
: Checking the floor mounting
: Check the spacing of the limit switches and the potentiomer play
: Checking installation of the linear guide
: Check the gear shaft of any cracks
: Clean and lubricate the tabletop, longitudinally and transversely
: Deburr the tabletop
: Tilt Drive - Checking the safety distance
Spotfilm Device and Column Longitudinal Drive - Checking the
:
Safety Distances
: Tabletop Longitudinal - Checking the Safety Distances
: Tabletop Transverse - Function Test
: SID - Function Test
: Perform a mechanical function test of safety devices
: Checking the unit Control Console for Function and Damage
Multileaf Collimator/Universal
V : Inspection and Maintenance
Collimator
: Checking the collimator mounting
: Checking the rotation flange (if configured)
: Checking the control elements on the collimator (if configured)
: Accessory rails under the collimator checked for wear
: Checking the light localizer lamp (if configured)
VI Monitor Carriage : Checking the mounting of the monitors
VII Wall Stand : Inspection and Maintenance
: Check and clean the contact surfaces
: Checking the Movement Range
: Checking the Braking Force
: Checking the Control Elements
: Checking of the Cassette Bucky
VIII Fluoro Detector : Check image quality detector
IX Final Work Steps : Contact voltage in volts
: Voltage phase 1 to the protective conductor in volts
: Voltage phase 2 to the protective conductor in volts
: Voltage phase 3 to the protective conductor in volts
X Function Test : Perform the final function check
2 Sparepart Bekas Tetap menjadi milik pabrikan Siemens
3 Lokasi Ruang Radiologi
4 Waktu kerja Senin-Jumat jam 09.00 s/d 22.00 WIB,
5 Hari Libur Sabtu, Minggu dan Hari Libur Nasional
Sesuai dengan Katalog Suku Cadang Siemens Healthineers, jika
6 Suku Cadang Service
terjadi kerusakan
7 Uptime Guarantee (%) Tidak Termasuk
8 Onsite Support Tidak terbatas jumlah kunjungan
9 Response Time Commitment Dalam 2 jam setelah panggilan
10 System Reliablity (%) Tidak Termasuk
11 Onsite Support (Upon Activation) Unlimited
12 Onsite Response Time (Emergency) 2 Hours
No. Nama Alat dan Uraian Pekerjaan Volume Merek
13 Onsite Response Time (Non - 2x24 jam
Emergency)