Uraian Pekerjaan
Pekerjaan Kontrak Service Pemeliharaan Alat Spect CT (Symbia 6/Syngo Via) Merek Siemens di Instalasi Radiologi
RSUP H. Adam Malik Tahun 2025
No. Nama Alat dan Uraian Pekerjaan Volume
1 Maintenance dan Kontrak Service Unit SPECT-CT Symbia Intevo 6 dan syngo.via: 1 Unit
Type Performance TOP
Specifications :
● Suku cadang Normal : Semua suku cadang yang ada pada unitSPECT-CT SymbiaIntevo 6dan
syngo.via
( Tidak Termasuk Radioactive Source, X-Ray Tube, Detector dan
:
Aksesoris )
● Preventive Maintenance (PM) : 2 (dua) kali selama 11 Bulan
● Corrective Maintenance (CM) :
Tidak terbatas (pada saat alat bermasalah atau mengalami kerusakan)
● Tenaga kerja dan perjalanan CM : Termasuk
● Tenaga kerja dan perjalanan PM : Termasuk
● tenaga kerja jarak jauh : Termasuk
● Dukungan telepon teknis : Termasuk
● Waktu respons awal : 2 (dua) jam
● Waktu respon di tempat (on-site) : Hari kerja berikutnya
● Waktu pengiriman suku cadang : 14 (empat belas) hari kerja
● Layanan jarak jauh teknisi (remote) : Remote pro-active support include
● Safety dan Update Software : Termasuk (jika ada instruksi dari pabrik)
● Jendela Pemeliharaan Perbaikan (CM : Setiap hari kerja dan jam kerja
● Jendela Pemeliharaan Direncanakan (PM) : Setiap hari kerja dan jam kerja
● Kalibrasi oleh instansi kalibrasi yang : Tidak Termasuk
berwenang dilengkapi dengan sertifikat, hasil uji serta
label kalibras
Preventive Maintenance Rutin, meliputi :
PM 1
General Visual Check : Visual Inspect System
- The system has no deficiencies
Cleaning the component : Cleaning System
1st - Maintenance : SRS Functionality Check
: SW Backup
: Trunnion Brakes
: Rotate Brakes
: PHS U/D Brakes
: SLD Acme Screw Inspection 1
: Control Box E-Stop for SPECT
: L/R E-Stop buttons for SPECT
: Control Box E-Stop for CT
: Radiation Indicators
: Radiation Shutdown
: Defragmentation on C: and D: Drive
: Swivel Caster Lubrication
: Front PHS Check
: Rear PHS Check
: Gantry tilt switch checked
: Caudal tilt detector - clutch connection (Detector 2) checked
Caudal tilt detector - pivot connection (Detector 2) checked
:
: Detector 1 Checked
Fixed detector yoke side to Tub Connection Detector 1 have been
:
tightened to reach torque spesification
: Fixed and Caudal tilt yoke arm connections checked
Fixed yoke side arm screws have been tightened to reach torque
:
spesification
Caudal tilt yoke side arm screws have been tightened to reach torque
:
spesification
: Yoke to trunnion checked
: Trunnion to SLD Connection checked
SLD to LLD connection checked
:
: Monitors, computers, and system cleaned
: No safety-relevant damages on the gantry
: Furniture/walls are over minimum distance from gantry
: PPM Monitor settings checked
: Motion disable on PPM when in dvd mode
: PPM mechanical operation checked
: No safety-relevant damages on the front PHS
: PHS floor plate screws are tight
: Rollers under the pallet checked
: Rolling force of pivot and its functionality checked
: ICC touchpads checked
: Light Rail touchpads checked
: No safety-relevant damages on the rear PHS
: Patient positioning aids checked, without visible damage
: swivel casters inspected
: L Arms checked
: ACC checked
: Cart present sensor checked
: Collimator cart checked
: SLD1 flapper checked
: Sensors at the back of yoke1 checked
: Foot pedal docker switch checked
: SPECT UPS cable checked
: SPECT UPS cleaned and checked
: Monitor settings checked
: CT Quality check / QA Constancy
Detector Testing intrinsic / extrinsic verification ( Tc99 source / Co57
:
source provide by hospital )
Syngo.via
: Check if cables are squeezed or broken
: Check all external cable connection
: Ensure that the plug are free of dust
: Safe cable routing
: Clean the monitor
: Clean the keyboard
: Check for minimum clearance
: Check the functionality of the on-site air conditioner
: Check the system for a clean environment
: Backup configuration and user-specific data
: Delete analyze file
: Check disk capacity
: Check licenses
: Check the messages files for error messages (e.g., HW problems)
: Check SRS connection
: Software backup performed
: Monitor image was checked (SMPTE image display)
: Protective conductor resistance test (< 200mOhm)
PM 2
General Visual Check : Visual Inspect System
- The system has no deficiencies
Cleaning the component : Cleaning System
2nd - Maintenance Gantry cooling fans checked
: Air filters checked
: Slip ring assembly cleaned
: Power brush assembly cleaned
: Power carbon brushes checked
: Signal data brushes checked
: Rotate limit sensors checked
: Gantry rotate operation checked
: Rotate ring lubricated
: Stiffening Ribs Checked
: Ring to base frame connections checked
: Motor connection checked
: Motor mount connection checked
Inspect for damager, loose parts, cracks/tears, screw connections
:
: Main bearing lubricated
: Tube cooler cleaned
: Checked drive belt tension checked
: Gantry brake checked
: Air filter checked
: PDS fan checked
: Surge suppressors checked
: Gantry Protective Conductor Resistance has been measured
: PHS: Protective conductor resistance has been measured
: LCB (Option) Protective conductor resistance has been measured
UPS (Symbia T/Intevo Series Systems): Protective conductor resistance
:
has been measured
ICS (Symbia T/Intevo Series Systems): Protective conductor resistance
:
has been measured
UPS (Symbia T/Intevo Series Systems): Protective conductor resistance
:
has been measured
IRS (Symbia T/Intevo Series Systems): Protective conductor resistance
:
has been measured
DRS (Symbia T/Intevo Series Systems): Protective conductor resistance
:
has been measured
: Monitor: Protective conductor resistance has been measured
Leakage Current for Patient Tabletop to system PE has been measured
:
(AC)
Leakage Current for Patient Tabletop to system PE has been measured
:
(DC)
Leakage Current for PPM Tabletop to system PE has been measured
:
(AC)
Leakage Current for PPM Tabletop to system PE has been measured
:
(DC)
Leakage Current for ECG Leads Tabletop to system PE has been
:
measured (AC)
Leakage Current for ECG Leads Tabletop to system PE has been
:
measured (DC)
Leakage Current for ECG Leads to Patient Tabletop has been measured
:
(AC)
Leakage Current for ECG Leads to Patient Tabletop has been measured
:
(DC)
Perform CT NM FoV Calibration (Tc99 source and Contrast Media
:
provide by hospital )
: CT Quality check / QA Constancy
Detector Testing intrinsic / extrinsic verification ( Tc99 source / Co57
:
source provide by hospital )
Syngo.via
: Check if cables are squeezed or broken
: Check all external cable connection
: Ensure that the plug are free of dust
: Safe cable routing
: Clean the monitor
: Clean the keyboard
: Check for minimum clearance
: Check the functionality of the on-site air conditioner
: Check the system for a clean environment
: Backup configuration and user-specific data
: Delete analyze file
: Check disk capacity
: Check licenses
: Check the messages files for error messages (e.g., HW problems)
: Check SRS connection
: Software backup performed
: Monitor image was checked (SMPTE image display)
: Protective conductor resistance test (< 200mOhm)
2 Sparepart Bekas Tetap menjadi milik pabrikan Siemens
3 Lokasi Ruang Radiologi
4 Waktu kerja Senin-Jumat jam 09.00 s/d 22.00 WIB,
5 Hari Libur Sabtu, Minggu dan Hari Libur Nasional
Sesuai dengan Katalog Suku Cadang Siemens Healthineers, jika
6 Suku Cadang Service
terjadi kerusakan
7 Uptime Guarantee (%) Tidak Termasuk
8 Onsite Support Tidak terbatas jumlah kunjungan
9 Response Time Commitment Dalam 2 jam setelah panggilan
10 System Reliablity (%) Tidak Termasuk
11 Onsite Support (Upon Activation) Unlimited
12 Onsite Response Time (Emergency) 2 Hours
13 Onsite Response Time (Non - Emergency) 2x24 jam
Merek
Siemens